Common Problem
?Q
What Hardware Protection Measures Must Be Followed Before Power-On and Cable Connection for the DCA4201 / DCA6201 / DCA1065 Sampling Oscilloscopes?
4、When connecting the TRIG cable, check the TRIG port for any damage. Before use, it is recommended to clean the RF connector with an alcohol-moistened cleaning swab.

?Q
SMU - device interface problem
?Q
SMU - electric shock damages the device under test
?Q
SMU - when the dual channel synchronization function is turned on, if the main channel is configured as a current source, connect the measured part (LR series element) to the probe of the main channel, and measure the magnetic coupling current of the measured part on the slave channel, if the: output1 command is not sent to the slave channel, only the: init command is sent to the slave channel, can the sensing current sampling data be read from the slave channel smoothly?
?Q
SMU - the command output1 precedes the command: will the issue of init affect the accuracy of measurement data? (difference in order of issue)
?Q
SMU - How should the master-slave configuration be set up for dual-channel synchronized sweeps on the S3022F? Do the two channels share a common ground internally?
When performing dual-channel synchronized operation on the S3022F, such as a DC pulse linear sweep, enable synchronization for both channels. Set the primary channel to Master and the secondary channel to Slave.
Common Ground: The two channels of the SMU do not share a common ground internally. If a common-ground measurement is required, the channels must be connected to a common ground externally.
?Q
SMU - is there a quantitative relationship between pulse width and pulse delay time?
?Q
SMU - trig: what are the trigger points set by count? If it is not set, what impact will it have on the test?
?Q
SMU - what is the concept of source delay time (MS), and What Happens If It Is Set Too Short?
Because the source output requires a certain amount of time to settle, the SMU needs a delay before taking a measurement. This allows the source output to settle before measurement, helping ensure more accurate results. This delay is called Source Delay.
The Source Delay should be longer than the source settling time. The required delay depends on factors such as the output amplitude and load characteristics. If the Source Delay is too short, the SMU may begin measuring before the source output has fully settled, potentially resulting in significant measurement error, with the measured value being lower than the actual value.