  • Q

    Why do some channels of single-mode VOA display locked mode and some channels display open-loop attenuation mode? How to switch modes?

    A
    There are three modes of single-mode VOA: power locking mode, open-loop attenuation mode and closed-loop attenuation mode
    Long press the mode switch button on the front panel to switch modes

    PS: refer to the operation manual for the meaning and operation of other front panel buttons


  • Q

    How to determine whether the single-mode VOA optical attenuator has power monitoring?

    A

    You can intuitively see whether there is input power & output power on the LCD UI panel, as follows:


  • Q

    Does the sampling oscilloscope dca4201 second generation host computer have real-time mode display?

    A
    Yes, select setup instument mode setup - in the GUI and check the tuning mode
  • Q

    Can the sampling oscilloscope display the equipment temperature on the upper computer of dca4201 second generation computer?

    A

    Yes, select setup display setup check temperature display in the GUI


  • Q

    How does the second generation of sampling oscilloscope dca4201 record the log log of the instructions sent by the automation software to the upper computer?

    A

    Select tools SCPI programming tools in the GUI and check it; The log will be saved to the log folder of the upper computer


  • Q

    How to set the threshold mode of the sampling oscilloscope dca4201 second generation computer when testing mask magin?

    A

    Mask test setting - select Maunal mode - set the corresponding threshold


  • Q

    What is the recommended light entry range of the second generation of sampling oscilloscope dca4201?

    A
    -About 3dB to - 6dbm
  • Q

    Is there light on the oscilloscope screen of the second generation of sampling oscilloscope dca4201?

    A
    First, confirm whether the device under test emits light, which can be checked with a power meter; Confirm that there is light on the power meter and no light on the oscilloscope. After the equipment is restarted and connected, conduct dark current calibration and check again; If none of the above works, confirm whether there is light of more than 0dB entering the optical port and damaging the internal devices
  • Q

    How does the sampling oscilloscope dca4201 display the current temperature of the device and the temperature during calibration on the GUI?

    A

    Measure configure measurement


Log in

Account number

Password

Register an account

Name

Please enter your name *

E-mail

Please enter your email address *

E-mail verfication code

Please enter your email verification code

Phone

Please enter your contact number

Password

Please enter your login password *

Confirm Password

Please enter your login password again *
Reset password

E-mail address

Please enter your email number *

E-mail verification code

Please enter your email verification code

New Password

Please enter your login password *

Confirm Password

Please enter your login password again *