    What Are the Test Power Ranges and Input Power Limits for the DCA6201 and DCA1065 Sampling Oscilloscopes?

    A

    The optical input power limits differ significantly between the two models. Strictly observe the following latest key specifications during testing:

    DCA6201 Specifications

    Test Sensitivity and Linear Measurement Range:

    • NRZ: -8 to 0 dBm
    • PAM4: -7 to -1 dBm
    Maximum Destructive Peak Optical Input Power: +7 dBm
    Exceeding this threshold will cause permanent damage to the optical input port.

    Maximum Trigger Input Peak-to-Peak Voltage: < 1.5 Vpp
    Do not apply an electrical signal exceeding this limit.

    DCA1065 Specifications

    Test Sensitivity and Linear Measurement Range:

    • PAM4: -5 to -1 dBm
    Maximum Destructive Peak Optical Input Power: +5 dBm

  • Q

    Why Is There No Signal in the Eye Diagram, or Why Does the Eye Diagram Look Abnormal on the DCA4201 Sampling Oscilloscope?

    A

    Check the following two items:

    1、Filter Setting

    Check whether the Filter is configured correctly and make sure it has not been mistakenly set to a data rate intended for an electrical port.

    2、External Signal

    Confirm that the external signal is normal by checking whether the CDR has successfully locked.

    If the eye diagram remains clearly abnormal and the issue is not related to the settings, please take a screenshot and contact FAE for further diagnosis.

  • Q

    Why Does Dark Current Calibration Keep Failing on the DCA4201 Sampling Oscilloscope?

    A

    Check the following three items:

    ① Confirm that no optical signal is entering the instrument.

    ② Confirm that the host computer is properly connected to the instrument.

    ③ Confirm that the instrument has been successfully switched to the correct data rate for the corresponding optical port.

    If the calibration still fails after confirming all the above, please contact FAE for further assistance.

  • Q

    How Should I Troubleshoot a Large Margin (Eye Diagram Margin) Measurement Deviation on the DCA4201 Sampling Oscilloscope?

    A

    Please check the following items in sequence:

    1、Mask Template Mismatch

    Confirm that the selected Mask template matches the current test data rate. The name of the active template is displayed in the lower-left corner of the eye diagram.

    2、Optical Input Power and Configuration

    Confirm that the optical input power is within the linear range of -8 to -2 dBm and that the correct wavelength is selected.

    3、Waveform Count

    Check whether the Waveforms setting is configured correctly.

    If the Margin measurement still shows a significant deviation after completing the above checks, perform a cross-check with another DCA under the same test conditions. After ruling out external environmental factors, contact FAE for further assistance.

  • Q

    What Could Cause an ER (Extinction Ratio) Measurement Deviation on the DCA4201 Sampling Oscilloscope?

    A

    There are two common causes:

    1、Excessive Temperature Difference (Dark Current Recalibration Required)

    If the current instrument temperature differs significantly from the temperature at which the previous dark current calibration was performed, the ER measurement may deviate. A temperature warning will appear below the eye diagram. It is recommended to set the temperature-difference warning threshold to 2–3°C under Calibration. If the warning appears, perform the dark current calibration again.

    2、Incorrect ERCF Setting

    Check whether the ERCF is correctly set for each data rate. The ERCF can be set to either a positive or negative value, but values beyond ±5% are not recommended.

    If the ER measurement still shows a deviation after completing the above checks, perform a cross-check with another DCA under the same test conditions. After ruling out external environmental factors, contact FAE for further assistance.

  • Q

    What Are the Strict Optical Input Power Limits for the DCA4201 Sampling Oscilloscope?

    A

    Strictly observe the optical input power limits to prevent overload damage to internal components.

    Optimal Linear Input Power Range: -8 dBm to -2 dBm (approximately -5 dBm recommended).

    Damage Threshold: Input power above +7 dBm will cause permanent damage to the instrument. Do not exceed this limit.

  • Q

    How Do I Troubleshoot a “Power Abnormal” Error or No-Light Indication When an Optical Signal Is Present on the DCA4201 / DCA6201 / DCA1065 Sampling Oscilloscopes?

    A

    Please troubleshoot the issue in the following three-step sequence:

    1、Optical Port Contamination

    Use a fiber inspection scope (FC/PC connector) to inspect the optical port. Contamination in the central optical transmission path may interfere with the signal and cause measurement errors. If contamination is found, clean it immediately with a fiber cleaning pen.

    2、Incorrect Setting

    Check whether the wavelength is set correctly in the host software. For example, if the signal under test is 1310 nm, make sure the wavelength has not been incorrectly set to 850 nm.

    3、Adapter Damage

    Visually inspect the fiber adapter in a well-lit environment. If the ceramic sleeve inside the adapter is cracked, replace the adapter.

    If all the above checks are completed and no issues are found, but the problem persists after restarting the instrument, please contact Lianxun FAE for confirmation before returning the instrument for repair.

  • Q

    How Do I Properly Connect the DCA4201 / DCA6201 / DCA1065 Sampling Oscilloscopes to a Host Computer?

    A
    All three models support three host-computer connection methods: USB, LAN, and ATE control.
    • USB Connection
    Click Setup → Connect Setup, select USB, and click Refresh. If the device serial number appears in the Serial Number field, communication has been established successfully.
    • LAN Connection

    The default device IP address is 192.168.170.7 (port 8807). Configure the local computer's IP address as 192.168.170.X, then select LAN to establish a direct connection.

    • ATE Control
    If the ATE software and the DCA GUI software are running on the same computer, set the IP address of the DCA to be controlled by the ATE to 127.0.0.1.
    If they are running on different computers, set the IP address of the DCA to be controlled by the ATE to the IP address of the computer running the DCA GUI software.
    The ATE communicates with the DCA through the DCA GUI. In both scenarios, the default port 8805 can be used.
  • Q

    Can Testing Begin Immediately After Powering On the DCA4201 / DCA6201 / DCA1065 Sampling Oscilloscopes?

    A
    No.
    Allow the instrument to warm up for at least 30 minutes before use. Once the internal temperature has stabilized, perform a dark current calibration.
    ⚠ Important: Dark current calibration must be performed under no-light conditions.
Log in

Account number

Password

Register an account

Name

Please enter your name *

E-mail

Please enter your email address *

E-mail verfication code

Please enter your email verification code

Phone

Please enter your contact number

Password

Please enter your login password *

Confirm Password

Please enter your login password again *

Authorization Key

Please enter your Authorization Key
Reset password

E-mail address

Please enter your email number *

E-mail verification code

Please enter your email verification code

New Password

Please enter your login password *

Confirm Password

Please enter your login password again *