High Precision Benchtop SMU
S2026H
Dual-channel High-precision SMU
Features

High Range
Range: ± 200V, ± 1A (DC), ± 3A (pulse)
High Resolution
The minimum measurement resolution can reach 10 fA/100 nV
High Sampling Rate
Supports up to 1M ADC sampling rate
Threshold Trigger
Hardware high-speed IO, capable of threshold triggering, enabling efficient interaction between output measurement values and user systemsFunctions and Advantages

DC I-V Output Capability

|
Voltage Programming Accuracy |
Range |
Measurement Resolution |
Accuracy (1 Year) ± (% reading+ offset)[1] |
Typical Noise (RMS) 0.1 Hz-10 Hz |
|
±200 V[2] |
100 μV |
0.03%+10 mV |
0.4 mV |
|
|
±20 V |
10 μV |
0.03%+1 mV |
50 μV |
|
|
±6 V |
1 μV |
0.03%+0.4 mV |
12 μV |
|
|
±0.6V |
100 nV |
0.03%+100 μV |
3 μV |
|
|
Temperature Coefficient |
±(0.15 × accuracy)/℃ (0℃-18℃, 28℃-50℃) |
|||
|
Maximum Output Power |
20W: ±20V@1A, ±200V@100mA; 20W: ±0.6 V@1A |
|||
|
Overshoot |
<±0.1% (typical. norma mode. step is 10 % to 90 % range, full range, resistive load) |
|||
|
Noise 10Hz-20MHz |
<5 mVrms (20V voltage source, 1A resistive load) |
|||
[1] Accuracy calculation example: To test the accuracy of a 600mV range with a 120mV output, the tolerance is:
[2] This instrument has a potentially dangerous high voltage (±210 V) output to the HI / Sense HI / Guard terminals. To prevent electric shock, relevant safety precautions must be taken before powering on. Do not connect the Guard terminal to any output, including shorting it to the chassis ground or output LO, as this will damage the instrument
|
Current Programming Accuracy
|
Range |
Measurement Resolution |
Accuracy (1 Year) ± (% reading+ offset) |
Typical Noise (RMS) 0.1 Hz-10 Hz |
|
±3 A[3] |
1 μA |
0.03% + 2 mA |
40 μA |
|
|
±1 A |
100 nA |
0.03%+90 μA |
7 μA |
|
|
±100 mA |
10 nA |
0.03%+9 μA |
600 nA |
|
|
±10 mA |
1 nA |
0.03%+900 nA |
60 nA |
|
|
±1 mA |
100 pA |
0.03%+90 nA |
6 nA |
|
|
±100 μA |
10 pA |
0.03%+9 nA |
700 pA |
|
|
±10 μA[4] |
1 pA |
0.03%+1 nA |
80 pA |
|
|
±1 μA[4] |
100 fA |
0.03%+200 pA |
20 pA |
|
|
±100 nA[4][5] |
100 fA |
0.06%+30 pA |
3 pA |
|
|
±10 nA[4][5] |
10 fA |
0.06%+9 pA |
600 fA |
|
|
Temperature Coefficient |
±(0.15 × accuracy)/℃ (0℃-18℃,28℃-50℃) |
|||
|
Maximum Output Power |
20W: ±20V@1A, ±200V@100mA; 18W: ±0.6 V@1A |
|||
|
Overshoot |
<±0.1% (typical. normal mode. step is 10% to 90% range, full range, resistive load) |
|||
[3] 3A range is available only for pulse mode, accuracy specifications for 3A range are typical
[4] For measuring weak small currents, it is recommended to continue using triple coaxial cables for connection. If the output method is changed from triple coaxial terminals to ordinary wiring, it will affect the current accuracy of the instrument
[5] Additional specification conditions: 10 PLC setting
|
Minimum Programmable Pulse Width |
100 μs |
|
Pulse Width Programming Resolution |
1 μs |
|
Pulse Width Programming Accuracy |
±10 μs |
|
Pulse Width Jitter |
2 μs |
|
Pulse Width Definition |
The time from 10 % leading to 90 % trailing edge as follows |

|
Item |
Maximums |
Maximum Pulse Width |
Maximum Duty Cycle |
|
1 |
0.1 A/200 V |
DC, no limit |
100% |
|
2 |
1 A/20 V |
DC, no limit |
100% |
|
3 |
3 A/66.6V |
1 ms |
5% |
|
4 |
3 A/160 V |
400 μs |
2% |
|
Source |
Maximum Output |
Typical Rise Time [6] |
Typical Settling Time[7] |
Test Load |
|
Voltage |
160 V |
800 μs |
1.2 ms |
NO load |
|
5 V |
40 μs |
100 μs |
NO load |
|
|
Current |
3 A~1 mA |
90 μs |
250 μs |
Full load[8] |
|
100 μA~10 μA |
120 μs |
400 μs |
Full load[8] |
|
|
1 μA |
800 μs |
1.2 ms |
Full load[8] |
|
|
100 nA |
2 ms |
5 ms |
Full load[8] |
|
|
10 nA |
5 ms |
20 ms |
Full load[8] |
[6] Leading edge, the time from 10% leading to 90% leading
[7] The time required from Pulse out 0 to reach within 1% of final value
[8] Test condition: Normal, resistive load 6V maximum output
|
Source |
Range |
Output settling time[9] |
Condition |
||
|
Fast[10] |
Normal |
Slow |
|||
|
Voltage |
200 V |
<500 μs |
<1 ms |
<2 ms |
Time required to reach within 0.1 % of final value at open load condition. |
|
20 V |
<60 μs |
<100 μs |
<600 μS |
||
|
6 V |
<60 μs |
<100 μs |
<300 μs |
||
|
0.6 V |
<50 μs |
<50 μs |
<50 μs |
||
|
Current |
3 A~1 mA |
<50 μs |
<100 μs |
<0.8 ms |
Time required to reach within 0.1 % (0.3 % for 3 A range) of final value at short condition. |
|
100 μA~10 μA |
<100 μs |
<150 μs |
<0.8 ms |
||
|
1 μA |
<1 ms |
<1 ms |
<1 ms |
||
|
100 nA |
<3 ms |
<3 ms |
<3 ms |
||
|
10 nA |
<10 ms |
<10 ms |
<10 ms |
||
[9] Output transition speed: Fast, Normal, Slow. Users can adjust the APFC parameters based on the load characteristics to obtain precision, and fast output characteristics
[10] Slow mode is recommended for overshoot sensitive equipment. Fast mode may have overshoot on output in some condition