PXIe Source Measure Unit
S2019C
4-channel PXIe SMU
The S2019C is a high precision, compact, cost-effective, 4-channel PXIe Source/Measure Unit (SMU) with the capability to source and measure both voltage and current. It has maximum ±40 V, ± 500 mA(DC), ±1A(Pulse) sourcing capability, and supports conventional SCPI commands for easy programming. It supports most standard PXIe chassis and multi-card synchronization. The S2019C can be integrated into production test environment to achieve test efficiency improvement and cost reduction.
Features

Easy to Expand
4-channel PXIe SMU, easy to scale out
High Range
Range: ± 40 V, ± 500 mA (DC), ±1A(pulse)
High Resolution
The minimum measurement resolution can reach 1 pA/1 μV
High Sampling Rate
Up to 1MS/s ADC sampling rateFunctions and Advantages

5 Functions In One Card
Voltage source

Can Test Various Equipment

Capture More Measurement Data
♦ 6.5-digit resolution: Enjoy best-in-class 6.5-digit sourcing and measurement resolution.
Rich Scanning Function
|
Voltage Accuracy |
Range |
Resolution |
Accuracy(1 year) ±(%RD+mV)[1] |
Typical Noise (RMS) 0.1 Hz-10 Hz |
|
±40 V[3] |
10μV |
0.015%+1.2 mV |
50 μV |
|
|
±6 V |
1 μV |
0.015%+600 μV |
10 μV |
|
|
Temperature Coefficient |
±(0.15×accuracy)/℃(0℃-18℃, 28℃-50℃) |
|||
|
Channel [2] |
CH0 to CH3 |
|||
|
Output Power |
4 W per channel and maximum 8 W total |
|||
|
Overshoot |
<±0.1%( typical, Normal, step is 10% to 90% range, full range, resistive load) |
|||
|
Noise 10Hz-20MHz |
6 V voltage source, 0.5 A resistive load, <3 mVrms |
|||
[1] Accuracy calculation example: Testing the accuracy of a 1V output in a 6V range, the
allowable tolerance is:
[2] All channel outputs are electrically isolated from the ground, but the outputs of each channel share a common ground(LO)
[3] This instrument has a potential danger of high voltage (±42 V) output to the HI/Sense HI/Guard terminals. To prevent electric shock, relevant safety precautions must be taken before powering on. Do not connect the Guard terminal to any output. (Including shorting to the chassis ground or outputLO),otherwise, the instrument will be damaged
|
Current Accuracy |
Range |
Resolution |
Accuracy(1 year) |
Typical Noise (RMS) 0.1Hz-10Hz |
|
±1 A[4][5] |
500 nA |
0.1% + 125 μA |
20 μA |
|
|
±500 mA[5] |
||||
|
±100 mA |
100 nA |
0.03% + 25 μA |
2 μA |
|
|
±10 mA |
10 nA |
0.03% + 2.5 μA |
200 nA |
|
|
±1 mA |
1 nA |
0.03% + 250 nA |
20 nA |
|
|
±100 μA |
100 pA |
0.03% + 25 nA |
2 nA |
|
|
±10 μA |
10 pA |
0.03% + 3 nA |
200 pA |
|
|
±1 μA |
1 pA |
0.03% + 500 pA |
30 pA |
|
|
Temperature Coefficient |
±(0.15×accuracy)/℃(0℃-18℃,28℃-50℃) |
|||
|
Channel |
CH0 to CH3 |
|||
|
Output Power |
4 W per channel and maximum 8 W total |
|||
|
Overshoot |
<±0.1%(typical. normal. step is 10% to 90% range, full range, resistive load) |
|||
[4] The 1A range only supports pulse mode, with a typical accuracy
[5] The 1A and 500mA ranges only support a voltage range of 6V