Standard Benchtop SMU
S3012H
Single-channel Precision SMU
The S3012H is compact and cost-effective benchtop Source/ Measure Units (SMUs) with the capability to source and measure both voltage and current. These capabilities make the S3012H ideal for a wide variety of IV (current versus voltage) measurement tasks that require both high resolution and accuracy.
Features

High Range
Range: ±200 V, ±3 A (DC), ±10 A (pulse)
High Resolution
Minimum measurement resolution up to 100 fA/100 nV
High Sample Rate
Up to 1M ADC sampling rate can be supported
Hardware Synchronization
Single machine and multi-machine synchronization, pure hardware high-speed synchronization, can achieve multi-channel low-latency synchronizationFunctions and Advantages

(5) Functions In One Body
Power source
♦ The first and third quadrants are the source: the actual polarity of the output V/I follows the source setting.
♦ The second and fourth quadrants are the load: CC and CV cooperate, when the load is used, the load setting polarity is opposite to the source polarity.

Various Devices Can Be Tested

Capture More Measurement Data
♦ 6½-digit digital resolution: the accuracy is equivalent to a 6½-digit digital multimeter.
Rich Scanning Functions
|
|
Range | Programming Resolution |
Accuracy (1 Year) ± (% reading+ offset) |
Typical Noise (RMS) 0.1 Hz-10Hz |
|||
| ±200 V | 1 mV | 0.02% + 30 mV | 1.5 mV | ||||
| ±20 V | 100 μV | 0.02% + 2 mV | 160 μV | ||||
| ±6 V | 50 μV | 0.02% + 500 μV | 36 μV | ||||
| ±200 mV | 1 μV | 0.02% + 120 μV | 4 μV | ||||
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) | ||||||
| Maximum Output Power | 30W: ±20V@1.5A; ±200V@0.1A; 18W: ±6 V@3A | ||||||
| Settling Time | <800 μs (typical) | ||||||
| Overshoot | <±0.1% (6 V voltage source, normal. step is 10% to 90% range, full range, esistive load) | ||||||
| Noise 10Hz-20MHz | <3 mVrms(6 V voltage source, 3 A resistive load) | ||||||
|
Current Programming Accuracy |
Range | Programming Resolution |
Accuracy (1 Year) ± (% reading+ offset) |
Typical Noise (RMS) 0.1 Hz-10 Hz |
|||
| ±10 A[1] | 50 μA | 0.4% + 40 mA | NA | ||||
| ±3 A | 15 μA | 0.05% + 2 mA | 40 μA | ||||
| ±1.5 A | 10 μA | 0.02% + 500 μA | 20 μA | ||||
| ±150 mA | 1 μA | 0.02% + 25 μA | 5 μA | ||||
| ±15 mA | 100 nA | 0.02% + 6 μA | 700 nA | ||||
| ±1.5 mA | 10 nA | 0.02% + 250 nA | 16 nA | ||||
| ±150 μA | 1 nA | 0.02% + 25 nA | 1 nA | ||||
| ±15 μA | 100 pA | 0.02% + 3 nA | 140 pA | ||||
| ±1.5 μA | 10 pA | 0.03% + 450 pA | 25 pA | ||||
| ±150 nA | 1 pA | 0.05% + 250 pA | 5 pA | ||||
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) | ||||||
| Maximum Output Power | 30W: ±20V@1.5A, ±200V@0.1A; 18W: ±6 V@3A | ||||||
| Settling Time | <500 μs (typical) | ||||||
| Overshoot | <±0.1% (6 V voltage source, normal. step is 10% to 90% range, full range, resistive load) | ||||||
[1] 10 A range is available only for pulse mode, accuracy specifications for 10 A range are typical
|
|
Range | Measurement Resolution | Accuracy (1 Year) ± (% reading+ offset) | ||||
| ±200 V | 100 μV | 0.02% + 30 mV | |||||
| ±20 V | 10 μV | 0.02% + 2 mV | |||||
| ±6 V | 1 μV | 0.02% + 500 μV | |||||
| ±200 mV | 100 nV | 0.02% + 120 μV | |||||
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃, 28℃-50℃) | ||||||
|
|
Range | Measurement Resolution | Accuracy (1 Year)± (% reading+ offset) | ||||
| ±10 A[1] | 10 μA | 0.4% + 25 mA | |||||
| ±3 A | 10 μA | 0.05% + 2 mA | |||||
| ±1.5 A | 1 μA | 0.02% + 500 μA | |||||
| ±150 mA | 100 nA | 0.02% + 25 μA | |||||
| ±15 mA | 10 nA | 0.02% + 6 μA | |||||
| ±1.5 mA | 1 nA | 0.02% + 250 nA | |||||
| ±150 μA | 100 pA | 0.02% + 25 nA | |||||
| ±15 μA | 10 pA | 0.02% +3 nA | |||||
| ±1.5 μA[2] | 1 pA | 0.03% + 450 pA | |||||
| ±150 nA[2] | 100 fA | 0.05% + 250 pA | |||||
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃, 28℃-50℃) | ||||||
|
|
Range | Measurement Resolution | Test Current | Typical Accuracy(1 Year)±(% reading+ offset) | |||
| 1 Ω | 1 μΩ | 1.5 A | 0.073% +0.3334 mΩ | ||||
| 10 Ω | 10 μΩ | 150 mA | 0.057% + 3.334 mΩ | ||||
| 100 Ω | 100 μΩ | 15 mA | 0.08% + 33.34 mΩ | ||||
| 1 kΩ | 1 mΩ | 1.5 mA | 0.057% + 333.4 mΩ | ||||
| 10 KΩ | 10 mΩ | 150 μA | 0.057% + 3.334 Ω | ||||
| 100 kΩ | 100 mΩ | 15 μA | 0.06% + 33.34 Ω | ||||
| 1 MΩ | 1 Ω | 1.5 μA | 0.06% + 333.4 Ω | ||||
| 10 MΩ | 10 Ω | 0.15 μA | 0.35% + 3.334 kΩ | ||||
| 100 MΩ | 100 Ω | 0.05 μA | 0.95% + 10 kΩ | ||||
|
Temperature Coefficient
|
±(0.15 × accuracy)/°C
(0℃-18℃,28℃-50℃) |
||||||
Source I mode, Manual Ohm Measurement (4-wire) | Total error = Vmeas/Isrc = R reading x (gain error % of V range + gain error % of I range + offset error of I source range/Isrc value %) + (offset error of V measure range/Isrc value) Example:I source value=1.5A at 1.5A range V measure range=6V range Total error (% reading + offset) = (0.02%+0.02%+500μA/1.5A+(500μV/1.5A) ≈ 0.073%+0.3334mΩ | ||||||
| Minimum Programmable Pulse Width | 100 μs | ||||||
| Pulse Width Programming Resolution | 1 μs | ||||||
| Pulse Width Programming Accuracy | ±10 μs | ||||||
| Pulse Width Jitter | 2 μs | ||||||
| Pulse Width Definition | The time from 10% leading to 90% trailing edge as follows |

| Item | Maximums | Maximum Pulse Width | Maximum Duty Cycle | ||||
| 1 | 0.15 A/200 V | DC, no limit | 1 | ||||
| 2 | 1.5 A/20 V | DC, no limit | 1 | ||||
| 3 | 3 A/6 V | DC, no limit | 1 | ||||
| 4 | 3 A/20 V | 1ms | 0.1 | ||||
| 5 | 10 A/6 V | 1ms | 0.1 |